Add to Cart
Millivolt Range and Accuracy | -10mV-75mV ± (O.OO5%+5uV) |
Resistance Range and Accuracy | 0Ω~400Ω ± (0.004%+0.002Ω) 200Ω-50KΩ ± (0.01/+0.5Ω) 50KΩ-300KΩ ± (0.03%) |
RTD Equivalent Temperature Accuracy (ITS-90) 4-wire probe | ±0.011℃@-100℃ ±0.015℃@-0℃ ±0.019℃@100℃ ±0.023℃@200℃ ±0.031℃@400℃ ±0.039℃@600℃ |
Thermistor Equivalent Temperature Accuracy | ±0.002℃@0℃ ±0.003℃@25℃ ±0.006℃@50℃ ±0.014℃@75℃ ±0.030℃@100℃ |
Thermocouple Equivalent Temperature Accuracy | ||||
Type | Range | Measurement Accuracy | ||
K | -200℃-0℃ 0℃-1370℃ | ±0.61℃ ±0.24℃ | ||
T | -200-0℃ 0℃-400℃ | ±0.60℃ ±0.25℃ | ||
S | -20℃-0℃ 0℃-500℃ 500℃-1750℃ | ±1.05℃ ±0.95℃ ±0.56℃ | ||
Accuracies are based on internal reference junction compensation. Please refer to Technical manual for equivalent accuracies with an external reference junction |